PPSU hip joint test, PPSU ball, PPSU knee joint test pad
Jelly pad, is a trial equipment for joints. It is used for joint surface replacement surgery. It is a necessary device in joint surgery operation. The joint test pad needs to be used in conjunction with the instruments such as metal models and testing. Joint flexion activity.
Different colors of PPSU can represent different specifications. Avoid repeatedly replacing test pads during surgery to achieve the purpose of simplifying surgery and shortening the operation time.
PPSU is a non-fixed-shaped, high-performance thermoplastic plastic. Compared with PSU and PEI, it has better impact resistance and chemical resistance.
After high-pressure steam sterilization failure test test, it proves that the PPSU has excellent hydrolysis and exceeds other non-fixed thermoplastic plastic plastic. In fact, PPSU can be placed in the steam sterilization environment unlimited, and high-pressure steam is widely used as a sterilization method for medical components, which makes PPSU a good material for medical equipment.
01
02
03
04
05
06
Color | Black | Natural color | Baby blue | Dark blue | Green | Yellow | Grey | Red | White |
Color number | BL001 | N000 | BB09 | DB0099 | GR350 | Y159 | GG6126 | R8040 | W002 |
![]() | ![]() | ![]() | ![]() | ![]() | ![]() | ![]() | ![]() | ![]() |
Polymer performance/type | Unit | PEEK5600G | PEEK5600GF30 | PEEK5600CF30 | PPSU | PSU | Test standard |
Fiber content | Not fill | 30%glass fiber | 30%carbon fiber | Not fill | Not fill | ---- | |
Tensile Strength | MPa | 101 | 163 | 220 | 72 | 69 | ISO 527-1/-2 |
Tensile modulus | GPa | 3.5 | 11 | 21 | 2.2 | 2.5 | ISO 527-1/-2 |
Elongation at break | % | 10-20 | 2.8 | 1.7 | 60-120 | 50-100 | ISO 527-1/-2 |
Bending strength | MPa | 157 | 260 | 321 | 91 | 103 | ISO 178 |
Flexural modulus | GPa | 3.5 | 10.5 | 17.9 | 2.4 | 2.6 | ISO 178 |
Notched impact strength | J/m | No Break | 860 | 637 | No Break | No Break | ISO 180/IU |
Heat distortion temperature 1.82MPa | ? | 160 | 315 | 315 | 210 | 174 | ISO 75-1/-2 |
Linear thermal expansion coefficient flow direction | ppm/? | 47 | 20 | 7 | 57 | 56 | ISO11359-1/-2 |
Linear coefficient of thermal expansion in the transverse direction | ppm/? | 50 | 67 | 56 | 57 | 56 | ISO11359-1/-2 |
Water absorption rate24h | % | 0.1 | 0.1 | 0.1 | 0.4 | 0.3 | ISO 62 |
Density | g/cm� | 1.3 | 1.53 | 1.41 | 1.29 | 1.24 | ISO 1183 |
ps:This data is the average value of the material and cannot be used as the acceptance standard. The average value of linear thermal expansion is between 0-150 ?.
With more than 18 years of experience in the application and development of polymer materials such as PEEK, Junhua Medical can support customers to complete one-stop services from raw material procurement to product OEM.